人抗凝血酶Ⅲ生产工艺中活性检测方法的建立及初步应用

  • 付艳丽 孔涛 郭玉婷 曹扬 梁凌宇
展开
  • 兰州生物制品研究所有限责任公司第四研究室,甘肃省疫苗工程技术研究中心,兰州 730046

网络出版日期: 2025-08-16

Establishment and preliminary application of active detection methods in human antithrombin Ⅲ production process

Expand
  • The Fourth Research Laboratory, Lanzhou Institute of Biological Products Co., Ltd., Vaccine Engineering Technology Center of Gansu, Lanzhou 730046, China

Online published: 2025-08-16

摘要

目的  建立和验证全自动凝血分析仪检测人抗凝血酶( human antithrombinⅢ,AT-)活性的方法。方法  参照全自动凝血分析仪操作说明书进行AT-Ⅲ活性的检测,并对方法的线性、检测范围、精密度和准确度进行方法学验证。结果  全自动凝血分析仪检测AT-Ⅲ活性的方法稳定,仪器自动运行标准曲线的决定系数均>0.980,定标血浆标准曲线线性范围为0.12~1.23 IU/ml,定标血浆高、中、低值和定量下限稀释样品的效价回收率均在80%~120%范围内,且高、中、低值效价回收率的相对标准偏差( relative standard deviation, RSD)<15%,定量下限稀释样品的效价回收率的RSD<20%,供试品的批内精密度RSD<5%,批间精密度RSD<10%,对3AT-Ⅲ制品及其中间品进行活性检测,每批样品分别测定3次的平均RSD<5%结论  全自动凝血分析仪检测AT-Ⅲ活性的方法具有较好的线性、检测范围、准确度和精密度,可用于AT-Ⅲ生产工艺中活性的检测。

本文引用格式

付艳丽 孔涛 郭玉婷 曹扬 梁凌宇 . 人抗凝血酶Ⅲ生产工艺中活性检测方法的建立及初步应用[J]. 国际生物制品学杂志, 2022 , 45(4) : 187 -190 . DOI: 10.3760/cma.j.cn 311962-20211111-00073

Abstract

Objective To establish and verify the method for detecting human antithrombin Ⅲ(AT-Ⅲ activity by automatic coagulation analyzer. Methods According to the operation manual ofautomatic coagulation analyzer, the AT-Ⅲ activity was detected, and the linearity and detection range, precision and accuracy of the method were verified. Results Automatic coagulation analyzer detectdAT-Ⅲ activity stably and automatically ran the standard curves with coefficients of determination all>0.980, the standard curve of the calibrator was 0. 12-1. 23 IU/ml, the titer recovery rate range of thehigh, middle, low value calibrator and quantitative lower limit dilution samples were 80%-120%.therelative standard deviations (RSDs) of the high, middle and low value titer recovery rates were<15%,and the RSD of quantitative lower limit dilution sample titer recovery rate was <20%, inter-batch RSDs for the test products were <5%, and the intra-batch RSDs were <10%. The average RSDs of 3 batches of AT Ⅲ products and intermediates tested 3 times were all <5%. Conclusion The method of automatic coagulation analyzer for detecting AT-Ⅲ activity has good linearity, test range, accuracy and precision, which can be used for the detection of AT-Ⅲ activity in the production process.

文章导航

/